Hil controller interface



Name of the equipment:

 Hil controller interface

 

Make & Model:

OPAL-RT Technologies, OP8660

 

I-Stem Registration ID-

……………………..                              

 

Category of Instrument

Characterisation & Testing/ Computational Facility

 

Types of Analysis / Testing

Electrical system requiring Hardware-in-loop study one sample at a time.

 

Application:

·         Data communication

·         signal conditioning

·         real-time processing

·         I/O connectivity

·         communication protocols

 

Description of Instrument

 

It is a Controller and data acquisition interface designed to be used with the complete simulation system to provide a supplementary signal conditioning.

Its core contains four OP5511 high current and high voltage input conditioning modules, which allow conversion of current and voltage to ±10V voltage signals.

 

Booking Details 

 

 

 


Book through I-STEM:

https://www.istem.gov.in/

 

Slot Booking Link

-

 

 

 

 

 

Booking available for

Internal and External Both

 

Requisition form for

Internals

Externals

Contact Details



Features, Working Principle and Specifications

Faculty In-charge:

 

Dr.Kumaravel s

Email ID:

kumaravels@nitc.ac.in

Phone number:

+91-9645654204

Technical Staff:

 

Sreeram pv -sreerampv@nitc.ac.in

+91-9961228949

 

Department

EED

 

Office Email ID

eedoffice@nitc.ac.in

 

Location

Room No. DB 202, Department of Electrical Engineering

 

Lab Phone No

0495 228 6301


Features of the equipment


Unique features/Measurement capabilities,        if any

Instrument Technical Description and Major Specifications (This Specifications Limited to Major 5)

 

 I/O connector: DB37F, DB9, banana jacks Operating temperature: 10 to 40 ºC (50 to 104ºF)

Storage temperature: -55 to 85ºC (-67 to 185ºF)

Input range: 15 A (factory default)

Signal output range: ± 10 Volts

Isolation: Galvanic, 2.5 kV

Bandwidth: DC to 100 kHz

Linearity: < 0.2%

Rise time: < 2 Microseconds

Power supplies: ±15 Volts

Measurement/Sample specifications:

Type of Sample Required for Analysis / Testing (Quantity, Pre-Preparation, State etc.) Guidelines for Sample Submission – User Instructions







Electrical system requiring Hardware-in-loop study one sample at a time.

User Charges Rs. (GST Extra)





Internal

External Academic Institutes

National R&D Lab

Industry

Nil

1000/-

2000/-

5000/-